V:Kit 5 GC & GCMS

green banner to illustrate page on v kit 5 gc qualification tests

V:Kit 5 GC includes software, a comprehensive protocol with over 30 user configurable GC qualification tests, traceable reference standards with certificates of Analysis and SOPS


V:Kit 5 GC qualification tests summary:

  • Inlets
    • Temperature accuracy & stability
    • Pressure accuracy & tightness
    • Split/purge gas flow accuracy & precision
  • Column ovens
    • Temperature accuracy & stability
  • Detectors
    • Temperature accuracy & stability
    • Detector gas flow accuracy & precision (H2, Air, Make-up)
    • Detector Response Linearity & precision (FID & TCD only)
    • Signal to noise and noise and drift – FID, TCD, ECD, NPD
  • Liquid autosamplers
    • Autosampler injection precision, carry-over and retention time reproducibility
  • Headspace Autosamplers
    • Precision and carry-over
    • Temperature accuracy & stability


GCMS

  • Detectors
    • RFPA test
    • Tune test – Mass Accuracy
    • Signal-to-noise test
    • Injector precision/Mass Ratio test
  • Column Oven
    • Temperature accuracy & stability
  • Inlets
    • Temperature accuracy & stability
    • Pressure accuracy & tightness
    • Split/Purge gas flow accuracy & precision
  • Liquid autosamplers
    • Autosampler injection precision, carry-over and retention time reproducibility
  • Headspace autosamplers
    • Precision carry-over
    • Temperature accuracy & stability