V:Kit Applications

V:Kit has the flexibility to qualify any equipment that has standard data associated with it.

Standard techniques for off-the-shelf V:Kit 5 techniques include:

Green icon to illustrate HPLC techniques, with link to hplc page
Green icon for GC and GCMS techniques, with link to more information on GC and gcms
green icon for V:Kit tablet dissolution, with link to more info on v kit for tablet dissolution
green icon to illustrate v kit 5 for uv vis
green v kit 5 icon to illustrate v kit 5 for analytical balances, no link yet.

photo of HPLC v-kit reference standards
HPLC V:KIT protocols include:

  • Pump flow rate accuracy and precision
  • Gradient composition accuracy and linearity
  • UV/Vis, PDA/DAD and fluorescence detector wavelength accuracy
  • UV/Vis, PDA/DAD, fluorescence, refractive index and ELS and conductivity response linearity
  • Autosampler injection volume precision and carry-over
  • Autosampler injection volume linearity
  • Column oven and autosampler tray temperature accuracy

photo for v-kit lcms qualification
LCMS V:KIT protocols include:

  • Mass Accuracy test
  • Signal to noise test
  • Pump Flowrate and Gradient proportioning test
  • Autosampler Precision, linearity & carryover test
  • Temperature Accuracy and Stability for Column and Sample Compartments

v-kit gcms protocols image
GCMS V:KIT protocols include:

  • Mass Accuracy test
  • Signal to noise test
  • Autosampler precision/mass ratio test
  • RFPA test (suitable for Agilent MS only)
  • Column oven, inlet and detector temperature accuracy & stability
  • Inlet pressure accuracy and tightness
  • Inlet and detector flow accuracy (inc. split, purge)
  • Autosampler injection precision, carryover and retention time reproducibility

v-kit GC protocols
GC V:KIT protocols include:

  • Column oven, inlet detector and detector temperature accuracy
  • Column oven, inlet and detector temperature stability
  • Inlet pressure accuracy and tightness
  • Inlet and detector flow accuracy (including split, purge, make-up, fuel gases
  • Autosampler injection precision, carryover and retention-time reproducibility
  • Detector response linearity (FID/TCD)

S5 Box