HPLC and GC service
GC v-kit software

GC v-kit version 3 software is included in the InitiaGC starter kit, follow the comprehensive SOPs to produce your qualifation reports. The software is independent of your Chromatography Data System (CDS). as there is no electronic data capture from your GC, v-kit software can be installed to any PC or laptop meeting the minimum requirements.

A 21 CFR part 11 compliant log-in module known as Administration Console is supplied with InitiaGC. This features user and administrator password protection, audit trail and pass expiry. Your system administrator has option of whether or not to use the compliance module at the time of installation.

One option is available for GC v-kit version 3:
summary reports - which enables electronic protocol and automatic summary report functions.


Step by Step: Use the GC v-kit version 3 software to log-in details of your chromatography equipment and your calibrated test equipment. You will be prompted to set test set-points and PASS/FAIL criteria - or simply use the pre-programmed default set-points and criteria. You will be guided through the process of raw data entry, data checking and reporting. The ultimate aim of the software is to generate an audit-ready report detailing the outcome of your qualification test.

GC v-kit Tests include:
• Temperature accuracy for oven, inlet and detector
• Temperature stability for column oven
• Pressure accuracy and tightness for inlet
• Flow rate accuracy and precision for inlet and detector
• Linearity and precision for complete GC system
• Injcetion precision for headspace sampler

 
The third generation of v-kit software is supplied with a declaration of software validation required for use in regulated environments
 


get a quote
 
View Example Report
 
 
 

Site Map | HPLC | GC | TOOLS | DownloadsCopyright © 2001 - 2007 "v-ki" v:kit" "vkit" nlg analytical ltd