v-kit Pro HPLC allows you to qualify your HPLC with UV/Vis, fluorescence, refractive index, evaporative light scattering and conductivity detectors using traceable standards, calibrated tools and controlled procedures.
 
  
 
 
 
 
 
 
 
 
 
 
The HPLC Inita kit is comprised of:
 
HPLC Protocol
 
Includes the following tests:

  • HPLC Flow Rate – pump A
    Uses a traceable liquid flowmeter to qualify an HPLC pump for flow rate accuracy and precision at 1, 2 or 3 flow rates
  • HPLC Flow Rate – pump B
    Uses a traceable liquid flowmeter to qualify a second HPLC pump or pump head for flow rate accuracy and precision at 1, 2 or 3 flow rates
  • HPLC Column Temperature
    Uses a traceable digital thermometer to qualify an HPLC column oven/cooler for temperature accuracy and precision at 1, 2 ,3 or 4 combination of temperature and position
  • HPLC Autosampler Temperature
    Uses a traceable digital thermometer to qualify a heated/cooled HPLC autosampler for temperature accuracy and precision at 1, 2 ,3 or 4 combination of temperature and position
  • HPLC Injection Precision
    This test uses a traceable precision standard and blank plus a qualified pump and UV/Vis, FL or, ELS or RI detector to qualify an HPLC autosampler for injection precision and carry over
  • HPLC UV Wavelength Accuracy
    Uses a traceable wavelength accuracy standard solution to qualify an HPLC UV/Vis detector for wavelength accuracy in manual, injection or scanned mode for 1, 2 or 3 wavelengths in the range 241 - 640nm
  • HPLC UV Extended Wavelength Accuracy
    Uses a traceable wavelength accuracy standard solution to qualify an HPLC UV/Vis detector for extended wavelength accuracy in manual, injection or scanned mode for 1 or 2 wavelengths, 205 and 273nm
  • HPLC FL Excitation Wavelength Accuracy
    Uses a traceable wavelength accuracy standard solution to qualify an HPLC fluorescence detector for excitation wavelength accuracy in manual or scanned mode at ex 350nm
  • HPLC FL Emission Wavelength Accuracy
    Uses a traceable wavelength accuracy standard solution to qualify an HPLC fluorescence detector for emission wavelength accuracy in manual or scanned mode at em 397nm
  • HPLC FL Extended Emission Wavelength Accuracy
    Uses a traceable wavelength accuracy standard solution to qualify an HPLC fluorescence detector for emission wavelength accuracy in manual or scanned mode at em 450nm
  • HPLC UV Signal–to–Noise or Noise and Drift
    Uses a traceable linearity standard to qualify an HPLC UV/Vis detector for signal noise, drift and signal-to-noise ratio or only signal noise and signal drift
  • HPLC FL Signal–to–Noise
    Qualifies an HPLC fluorescence detector for signal-to-noise ratio
  • HPLC RI Signal–to–Noise or Noise and Drift
    Uses a traceable linearity standard to qualify an HPLC refractive index detector for signal noise, drift and signal-to-noise ratio or only signal noise and signal drift
  • HPLC ELS Signal–to–Noise or Noise and Drift
    Uses a traceable linearity standard to qualify an HPLC evaporative light scattering detector for signal noise, drift and signal-to-noise ratio or only signal noise and signal
    drift
  • HPLC UV Linearity
    Uses five traceable linearity standards plus a qualified pump and autosampler to qualify an HPLC UV/Vis detector for response linearity
  • HPLC FL Linearity
    Uses five traceable linearity standards plus a qualified pump and autosampler to qualify an HPLC fluorescence detector for response linearity
  • HPLC RI Linearity
    Uses five traceable linearity standards plus a qualified pump and autosampler to qualify an HPLC refractive index detector for response linearity
  • HPLC Injection Linearity
    Uses a traceable precision standard plus a qualified pump and UV/Vis, FL or RI detector to qualify an HPLC autosampler for injection linearity
  • HPLC Gradient Accuracy – A/B
    Uses a traceable gradient standard and a qualified UV/Vis or FL detector to qualify an HPLC pump for gradient accuracy and at 5, 6 or 11 composition steps
  • HPLC Gradient Accuracy – C/D
    Uses a traceable gradient standard and a qualified UV/Vis or FL detector to qualify an HPLC pump for gradient accuracy and at 5, 6 or 11 composition steps

HPLC technique license
Either pre-loaded on the Dongle or supplied as a V2C file the technique license allows the user to start a protocol.
 
HPLC accessories
1 x pressure loop, 2 x fingertight fittings, 1 x ZDV union, 1 x luer adaptor, 1 x 20mm de-crimper, 1 x Polypropylene syringe & needle
 
 
 
Click here to order the v-kit Pro HPLC Initia Kit and associated Refill Kits
 
Click on the image below to download the v-kit Pro HPLC datasheet
 
Basket
British Pound - GBP
Your basket is empty.

Add using Quick Order...



v-kit is operated by NLG Analytical Ltd, 16 John Bradshaw Court, Alexandria Way, Congleton Business Park, Congleton, Cheshire, CW12 1LB, UK

Tel: +44(0)1260 274-266 Fax: +44(0)1260 273-319

© 2012 NLG Analytical Ltd. All rights reserved.

sectionid=595&urltype=7&sssectionid=595&totype=ARTICLE&toid=89&pagetemplateid=5&pagetemplatecode=&articleid=89&language=en&sssectionname=v-kit+Pro+HPLC&brandid=1
Page as at 19 May 2012 16:07:22.